70
|
ERROR MESSAGES
Message
Description
Aborted
Indicates a cycle was manually aborted and not completed
Air Disabled
Indicates that the air flow sensor was disabled
A Dose Error
Alarm indicating that the SSG Part A dosing reservoir has timed out and not all
chemistry was delivered
Bas Disabled
Indicates that the basin level sensor was disabled
Bas Sen Err
Indicates that the basin level sensor is reporting a full basin at the start of a cycle (basin
should be empty)
Basin Temp
Indicates that the basin temperature has not reached the programmed minimum
B Dose Error
Alarm indicating that the SSG Part B dosing reservoir has timed out and not all
chemistry was delivered
Clear Sys Alarm?
Diagnostic 70: prompts the user to press the Enter button to clear a system alarm
(such as a NVRAM error)
Flash High Error
Indicates that the high code flash CRC test did not pass
Flash Low Error
Indicates that the low code flash CRC test did not pass
Flo Disabled
Indicates that the fluid flow sensor was disabled
Flow Sen Err
Indicates that the flow sensor is reporting flow at the start of a cycle (no flow should be
occurring)
Flow 2 Sen Err
(SSG only) alarm indicating that the chamber fluid flow sensor is reporting flow at the
start of the cycle (no flow should be occurring)
HLD Sampl Er
Alarm indicating that a disinfectant sample port sensor is reading incorrectly
Lid ajar
Indicates that the cover (lid) is open during a cycle
Lid Disabled
Indicates that the cover (lid) open sensor was disabled
Low Alcohol
Indicates that the alcohol level is low
Low Chamber
Indicates that the basin level did not reach the level sensor in time