71
Message
Description
Low Det.
Indicates that the detergent level is low
No Air Flow
Indicates that no air flow was detected
No Fluid Flo
Indicates that no fluid flow was detected
No Fluid Fl2
(SSG only) alarm indicating that no fluid flow was detected by the 3.2 l/min chamber
fluid sensor
NVRAM Err
Non-Volatile RAM error, can only be cleared in Diagnostics 70
Power on
Indicates that the device lost power during a cycle
RAM Error
RAM error, can only be cleared in Diagnostics 70
Reset Alarms?
Prompts the user to press the Enter button to cancel an alarm
Leak Disabled
Indicates that the leak tester was disabled
Leak Fail
Indicates that the leak test failed
Leak Sen Err
Indicates that the leak tester measured pressure at the start of a cycle
SNVRAM Err
Short Non-Volatile RAM error, can only be cleared in Diagnostics 70
SSG A low
Alarm indicating that the SSG has timed out before filling part A reservoir
(Part A chemistry low)
SSG B low
Alarm indicating that the SSG has timed out before filling part B reservoir
(Part B chemistry low)
SSG A Sen Er
Alarm indicating that the SSG dosing reservoir A sensors are reading incorrectly
(sensor hardware error)
SSG B Sen Er
Alarm indicating that the SSG dosing reservoir B sensors are reading incorrectly
(sensor hardware error)
SSG Cmd Err
Alarm indicating that the SSG has received an unknown command from the DSD
(software error?)
SSG Comm Err
Alarm indicating that the SSG detected a failure in the DSD Communications
SSG CRC Fail
Alarm indicating that the SSG failed its internal Software Checksum test (program not
functioning correctly)