55
OPEN INTERNAL WATER SUPPLY VALVE
This function is typically used as part of procedures for disinfecting water lines and changing water filters.
1. On the control panel, press
SETUP
2. Type
43
on the keypad, then press
ENTER
.
This opens the incoming water supply solenoid valve providing a water supply to the on-board .2 micron
filter housing.
A counter on the LCD display begin to count down from 250 to 0 in one second increments. After 250 seconds,
the incoming water supply solenoid valve automatically closes.
3. To manually close the internal water supply valve, press
STOP
.
4. To exit, press
SETUP
.
CREATE A CUSTOM CYCLE PROGRAM
Custom programs allow the operator to change the cycle parameter settings, or to set up custom reprocessing
protocols. A maximum of nine custom programs can be pre-set.
Note:
Depending on selections, some of the following screens may not display.
Note:
To deactivate a cycle, type
00
for the number of minutes and seconds.
Note:
Press
STOP
at any time to exit this function.
Service:
For assistance with identifying and resolving the issue,
contact Medivators Technical Support at 1-800-444-4729.