43
CONTINUE A CYCLE AFTER AN OPERATOR INTERRUPTION
1. On the control panel, verify the correct station is selected. If necessary, press
STATION SELECT
to choose
the station (A or B).
2. Press
START
. The cycle proceeds as normal.
TERMINATE A CYCLE AFTER AN INTERRUPTION
1. On the control panel, verify the correct station is selected. If necessary, press
STATION SELECT
to choose
the station (A or B).
2. Press
CANCEL
.
3. Press
ENTER
.
“Aborted” displays on the screen. The cycle has been successfully terminated. When a cycle is terminated,
the reprocessor fails to a safe mode and the disinfection cycle must be re-run for the endoscope.
An endoscope must not be used until the entire disinfection process has successfully completed.
SHUTDOWN THE REPROCESSOR
The reprocessor must be properly shut down at the end of each day.
1. Close the main (external) water supply valve.
2. Sanitize the reprocessor upper basins and basin lids
with an EPA-registered sanitizer, such as properly diluted
ACTRIL™ Cold Sterilant. Follow the sanitizer manufacturer’s
recommendations for proper use.
Figure 5.5
- Close External Water
Supply Valve
Warning!
Avoid possible chemical burns. Always wear personal protective
equipment (gloves, goggles) when handling sanitizer.