30
|
CHECK DISINFECTANT CONTAINERS
Disinfectant is supplied in two separate 5-liter containers. Part A is the Peracetic Acid component, Part B is the
buffer. Both containers fit into the disinfectant drawer located on the left-hand side of the reprocessor.
Warning!
Avoid possible chemical burns. Always wear personal protective
equipment (gloves, goggles) when handling disinfectant.
Warning!
Avoid possible slip injuries. Clean up any spills immediately.
Note:
On the control panel LCD screen, error message “Low Part A, Low Part B” displays if
a disinfectant container is low and needs to be replaced. Do not manually determine whether
disinfectant levels are low.
Note:
If Part A or Part B need to be replaced, replace both containers. Always replace both
containers at the same time.
Note:
Part A Uptake Tube is fitted with the BLUE cap. Part B Uptake Tube is fitted with the
WHITE cap.
Note:
Ensure that Part A and B caps are securely screwed onto containers.
1. Open the reprocessor doors and slide the
disinfectant drawer fully out.
Figure 4.4.
- Disinfectant Drawer