41
4. Open the reprocessor lid.
5. Remove the basin floating lid.
6. Verify the hookups are securely connected to the endoscope.
7. Disconnect the hookup connectors from the endoscope.
8. Remove the endoscope from the basin.
Warning!
If any scope hookup connections were disconnected during the cycle,
the endoscope was not properly disinfected and is not patient ready. Re-run the
disinfection cycle for the endoscope to ensure the endoscope is properly disinfected
and rinsed.
HOW TO PROCEED WHEN A LEAK IS DETECTED
Caution!
A leak test must be performed using a manual leak test or validated Dry Leak Test
System prior to reprocessing in the DSD EDGE™ Reprocessor. The Leak Testing system in
the DSD EDGE Reprocessor is for redundancy purposes to protect the integrity of the scopes.
Major Leak Detected
If the pressure drops below 50 mmHg, the cycle is halted,
the control panel’s Warning LED indicator blinks and a
“Leak Fail” error message displays on the screen.
1. On the Control Panel, press
STOP
to end
the cycle. The cycle is aborted.
2. Resolve the leak issue.
Figure 5.4
- Leak Fail Message
Service:
For assistance with identifying and resolving the issue,
contact Medivators Technical Support at 1-800-444-4729.
3. Once the issue has been resolved, start a new disinfection cycle.