42
|
Minor Leak Detected
If a minor leak was detected, the cycle continues until completion. At the end of the cycle, the control panel’s
Warning LED indicator blinks and a “Leak Fail” error message displays on the screen.
1. On the Control Panel, press
STOP
to clear the warning message.
2. Resolve the leak issue.
Service:
For assistance with identifying and resolving the issue,
contact Medivators Technical Support at 1-800-444-4729.
HOW TO RESOLVE A PROCESS INTERRUPTION
A process interruption may have been initiated by the operator or result from a system interruption. A system
interruption may be caused by loss of water or air, loss of power, or by a lack of available Part A and Part B
disinfectant components.
Note:
If the interruption is caused by a power outage, the cycle automatically restarts when
power is restored. A “Power On” message displays in the log.
CONTINUE A CYCLE AFTER A SYSTEM INTERRUPTION
1. Resolve the issue causing the error.
Service:
For assistance with identifying and resolving the issue,
contact Medivators Technical Support at 1-800-444-4729.
2. Once the issue is resolved, press
START
on the control panel. The reprocessor continues the cycle from the
point of interruption.