Page 35
Sonic-Pro
Industries, Ltd.
Industries, Ltd.
ProSeries
by Blue-White Ind.
TM
1.
Press the
SETUP
button.
yellow
yellow
he
meter will re-start and enter the run
mode. While the meter is restarting, the
four yellow indicator lights will go out,
one by one, as shown in step 3 above.
After the meter re-starts, the yellow
indicator lights will go out. Setup is
complete.
The two outer
indicator lights will blink,
indicating that the meter has entered
the setup state.
2.
Press the
SETUP
button again. The two
inner
indicator lights will blink.,
indicating that the meter has entered
the transducer positioning state. Adjust
the transducers if necessary. Press the
SETUP
button when this is complete.
3.
The meter will re-start. While the meter
is restarting, the four yellow indicator
lights will go out, one by one. During
this time, the zero flow calibration
should be performed. To enter the zero
calibration state, press and hold the
SETUP
button for at least 3 seconds
and then release.
4.
The meter will perform a zero flow
calibration. One yellow indicator light
will be lit.
Be sure there is no
movement of fluid in the pipe
during the zero calibration
.
5.
After a successful zero calibration, t
6.
The yellow lights will return once flow
measurement begins. During flow
measurement, the lights will indicate
the Goodness of Measurement index:
one light = poor quality
two lights = good quality
three lights = getter quality
four lights = best quality
INDICATOR LIGHTS IN
SETUP STATE
INDICATOR LIGHTS IN
TRANSDUCER POSITIONING STATE
INDICATOR LIGHTS IN
STARTUP STATE
INDICATOR LIGHTS DURING
ZERO FLOW CALIBRATION
INDICATOR LIGHTS AFTER
START-UP STATE
5.7
Model S1
Zero Flow
Calibration
INDICATOR LIGHTS DISPLAYING
GOODNESS OF MEASUREMENT