Page 33
Sonic-Pro
Industries, Ltd.
Industries, Ltd.
ProSeries
by Blue-White Ind.
TM
2.
The display will indicate the currently
configured transducer separation
distance and mounting method. Adjust
the transducers if necessary. Press the
SETUP button when this is complete.
3.
The meter will re-start. During this
time, the zero flow calibration should be
performed. While the meter is
restarting, press and hold the
SETUP
button for at least 3 seconds and then
release.
4.
The meter will perform a zero flow
calibration.
Be sure there is no
movement of fluid in the pipe
during the zero calibration
.
5.
After a successful zero calibration, the
run screen will appear. Setup is
complete.
SETUP
Tap Setup to enter Setup
Hold Setup for Zero Calibration
Startup …
[
£
£
£
£
¢
¢
¢
]
Information Screen
Configuration Set #1
Blue-White Industries, Sonic-Pro
Firmware Version, A01.00.00
(Searching for SD Card)
SETUP
Zero Calibration …
[
£
£
£
£
¢
¢
¢
]
Information Screen
Blue-White Industries, Sonic-Pro
Firmware Version, A01.00.00
¢
¢
¢
1224
SETUP
SWAP
CLEAR
U.S. Gallons
1 T 2.48 in.
GM:[
¢
¢
¢
£
]
U.S. Gallons / min
625.3
0
SETUP
SWAP
CLEAR
U.S. Gallons
1 T 2.48 in.
[
¢
¢
¢
£
]
U.S. Gallons / min
0.0
Position the transducers
at a separation distance of:
Transducer Positioning …
2.489416 inches (”V” Mount)
Blue-White Industries, Sonic-Pro
Firmware Version, A01.00.00
Press DONE when this is complete
Information Screen
DONE
5.5
Model S2
Zero Flow
Calibration