Run Mode
, the flowmeter performs flow measurement, updates the 4 to 20 mA
and pulse signal outputs, displays and logs data and performs process control actions
according to the active configuration.
The flowmeter indicates that it is in the Run Mode as follows:
The flowmeter remains in the
Run Mode
until it is powered down, the user takes an
action to go to Setup Mode or, exceptionally, an error occurs and the flowmeter goes to
the
Faulted State
. The user can command the flowmeter to
Setup Mode
by pressing
the motherboard setup button or changing the position of the measurement mode switch
between Transit Time and Doppler. If the
Model S3 Display
or the
Communications
PC Software
is connected, the user can additionally press the
SETUP
soft button.
For a complete description of
Run Mode
operation, see section 7 below.
2.9
Run
Mode
Model S1:
Yellow LEDs indicate the
present “Goodness of Measurement”.
one light = poor quality
four lights = best quality
Model S2:
as for Model S1, plus the
Run
Mode
screen appears, including an
indication of “Goodness of Measurement” on
the top line and the calculated speed of
sound in the fluid if in Transit Time
measurement mode.
Model S3:
as for Model S2 (see section 7 for
some minor differences between Model S2
and S3 displays).
MODEL S2 and MODEL S3