![Blue-White ProSeries Sonic-Pro S1 Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/blue-white/proseries-sonic-pro-s1/proseries-sonic-pro-s1_installation-and-operating-manual_2776995031.webp)
Page 31
Sonic-Pro
Industries, Ltd.
Industries, Ltd.
ProSeries
by Blue-White Ind.
TM
0
SETUP
SWAP
CLEAR
U.S. Gallons
1 T 2.48 in.
[
£
£
£
£
]
U.S. Gallons / min
ENTER
Ç
ESC
È
Setup Root Menu
Manage Configuration Sets
Global Configuration
SETUP
Zero Calibration …
[
£
£
£
£
¢
¢
¢
]
Information Screen
Blue-White Industries, Sonic-Pro
Firmware Version, A00.05.02
¢
¢
¢
1.
Model S3 includes a full featured
display with 5 button touch pad. Press
the
SETUP
button.
2. The meter will enter the
SETUP ROOT
MENU
.
Creation and modification of
the five available Configuration Sets
and the Global Configuration settings
can be made from this menu. Refer to
section 6 on page 36 for details. If no
changes are to be made, press
ESC
.
3.
The meter will re-start. During this
time, the zero flow calibration function
can be performed. While the meter is
restarting, press and hold the
SETUP
button for at least 3 seconds and then
release.
4.
The meter will perform a zero flow
calibration.
Be sure there is no
movement of fluid in the pipe
during the zero calibration
.
5.
After a successful zero calibration, the
run screen will appear. Setup is
complete.
SETUP
Tap Setup to enter Setup
Hold Setup for Zero Calibration
Startup …
[
£
£
£
£
¢
¢
¢
]
Information Screen
Configuration Set #1
Blue-White Industries, Sonic-Pro
Firmware Version, A00.05.02
(Searching for SD Card)
1224
SETUP
SWAP
CLEAR
U.S. Gallons
1 T 2.48 in.
GM:[
¢
¢
¢
£
]
U.S. Gallons / min
625.3
0
SETUP
SWAP
CLEAR
U.S. Gallons
1 T 2.48 in.
[
¢
¢
¢
£
]
U.S. Gallons / min
0.0
5.2
Model S3
System
Start-Up
5.3
Model S3
Zero Flow
Calibration