Page 15
Sonic-Pro
Industries, Ltd.
Industries, Ltd.
ProSeries
by Blue-White Ind.
TM
2.8
Transducer
Positioning
State
In the
Transducer Positioning State
, the flowmeter waits for the transducers to be
correctly positioned on the pipe.
The flowmeter indicates that it is in the
Transducer Positioning State
as follows:
You must acknowledge that the transducers are correctly positioned by either pressing
the
SETUP
button on the motherboard or the
DONE
soft button on the display, if fitted.
The flowmeter will then return to the
Startup
state, see section 2.5 above, from
which you can again enter
Setup Mode
or perform
Zero Calibration
or, by allowing
the 7 second timeout to elapse, let the flowmeter enter
Run Mode
and begin flow
measurement.
When operated in the Transit Time measurement mode, the separation distance of the
transducers is computed by the flowmeter from the information in the
Transducer
Setup
branch of the
Configuration Menu
. The correct separation distance is
dependent on the pipe OD, pipe wall thickness, pipe type and chosen mounting mode. If
the Doppler measurement mode is selected (by the motherboard switch), the separation
of the transducers (along the pipe axis) is always zero.
When factory configured, the separation distance and mounting mode is printed on the
serial label. The separation distance will also be displayed on the LCD display or on the
optional communications PC software, in either inches or millimeters depending on the
Measurement Units
chosen in
Transducer Setup
. If the meter has not been factory
configured, or if a new configuration is required, the pipe OD, pipe wall thickness, pipe
type and chosen mounting mode configuration data must be input before proceeding. See
section 6 for instructions on how to create or modify a configuration.
If the flowmeter has neither an LCD display nor a connected user
PC Software
application, then the Configuration information cannot be changed by the user so the
separation of the transducers will not change from the factory setting.
Model S1:
two center yellow LEDs flash at
1 Hz.
Model S2:
as for Model S1, plus the screen
presents instructions to the user as shown.
Model S3:
as for Model S2.
TRANSDUCER
POSITIONING STATE
DONE
Position the transducers
at a separation distance of:
Transducer Positioning …
Information Screen
2.489416 inches (”W” Mount)
Blue-White Industries, Sonic-Pro
Firmware Version, A01.00.00
Press DONE when this is complete