Page 22
Sonic-Pro
Industries, Ltd.
Industries, Ltd.
ProSeries
by Blue-White Ind.
TM
4.1
Transducer
Piping System
Location
NO
NO
Air bubble
Sediment
OK
OK
OK
OK
NO
NO
Down flows
must have
back pressure
Pipe must be full
Air could be trapped
!
10 x D
5 x D
Flow
Direction
4.3
Transducer
Mounting Mode
for
Transit Time
Measurement
TRANSIT TIME MOUNTING MODES
DOPPLER MOUNTING MODE
All pipe types and sizes
Select a pipe location that provides a minimum straight length of pipe of at least 10
times the pipe’s nominal pipe size and mark a point that is at least 5 times the pipe
diameter downstream from the nearest pipe fitting. See section 1.5 for the minimum
straight pipe length requirements.
If the fluid to be measured contains more than 10% particles, the meter should be
operated using the Doppler measurement method. For Doppler operation, the
transducers will be mounted directly opposite each other as shown below.
If the fluid contains little or no particles (up to 10% maximum), operate the meter using
the Transit Time method to obtain the best accuracy. Select a mounting mode that allows
the greatest sound travel distance while permitting good signal strength. Basic pipe
material and size guidelines are shown below.
TRANSDUCER MOUNTING LOCATION
4.0 Transducer Installation
4.2
Transducer
Mounting Mode
for
Doppler
Measurement
2” to 6”
NA
“V” Mount Mode
Pipe Material
Pipe Size
Plastic
Metal
Pipe Material
Pipe Size
6” to 100”
2” to 100”
“Z” Mount Mode
Pipe Material
Pipe Size
Plastic
Metal