Air bubble
Sediment
NO
NO
Down flows
must have
back pressure
Pipe must be full
Air could be trapped
!
Page 23
Sonic-Pro
Industries, Ltd.
Industries, Ltd.
ProSeries
by Blue-White Ind.
TM
When the meter is operated in the
Doppler
measurement method, the transducers are
ALWAYS mounted directly opposite each other. The separation is therefore zero.
When the meter is operated in the
Transit-Time
measurement method, the front faces
of the transducers must be positioned the correct distance apart.
The correct separation distance is dependent on the pipe size, pipe type and chosen
mounting mode. When factory configured, the separation distance and mounting mode is
printed on the serial label. If the meter has not been factory configured, or if a new
configuration is required, the
configuration data
must be input before proceeding. See section 6 for instructions on how to create or modify
a configuration.
For meters so equipped, the separation distance (as calculated based on the currently
active configuration set) is also shown on the LCD display and on the Remote
Communications PC Software display screen.
pipe size, pipe type and mounting mode
4.4
Transducer
Separation
Distance
Separation
Distance
625.38
4345625.38
SETUP
SWAP
RELAY
CLEAR
U.S. Gallons
Relay #2
Relay #1
off
Active
5 T 2.327 in.
[
¢
¢
¢
£
]
U.S. Gallons / min
Currently Active Configuration set
Separation Distance
Currently Active Configuration set
Separation Distance
LCD Display Screen
Communications PC Software Screen
A
B
1480