Use only Bionet-designated sensors. Other sensors may not provide adequate
protection against defibrillation or may put the patient at risk.
Disposable accessories (disposable electrodes, transducers, etc.) should be used only
once. Do not reuse disposable accessories.
SpO2 Patient Preparation
The accuracy of SpO2 monitoring is largely dependent on the strength and quality of the SpO2
signal.
If you use your fingers as a monitoring site, remove the nail polish. Cut the patient's fingernail if
needed to improve placement of the sensor. Only use sensors provided by Bionet and apply them
according to manufacturer's recommendations on a per-sensor basis.
If the sensor is not attached correctly, the ambient light may interfere with the pulse oximetry,
making the measurement irregular or causing the value to disappear. If you suspect interference
from ambient light, make sure that the sensor is properly positioned and that the sensor cover with
the opaque body is covered.
Preparation Order
1.
Select the sensor type and size that best suits your patient.
2.
If the sensor can be reused, please wash it before use for each patient.
3.
Position the sensor correctly and attach it to the patient.
4.
Connect the sensor to the patient cable.
5.
Check the application area of the sensor from time to time. If the sensor is too tight, it
may delay blood flow or overheat the skin and damage the tissue. Do not use a damaged
sensor.
Содержание BM1
Страница 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Страница 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Страница 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Страница 55: ...Rev 2 2 55 ...
Страница 62: ...Rev 2 2 62 Alarm Review ...
Страница 66: ...Rev 2 2 66 Click the numeric window you want to edit Edit total score items Edit numeric items Edit text item ...
Страница 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Страница 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...