Network for medical devices in hospitals
Network for portable patient monitors
When Using an Inappropriate Network
If your network does not meet the requirements, the following dangerous situations can occur.
The following situations may occur with this unit.
If the distributed alarm system is not safe:
-
The alarm will not be delivered.
-
The alarm or data is delayed.
-
An error alarm appears.
If the network connection is interrupted:
-
The alarm will not be delivered.
-
Reactivates with the alarm off or the alarm sound off.
If you do not have firewall and antivirus software:
-
Your data is not protected.
-
The device settings are changed.
-
The device raises an error alarm or does not generate an alarm.
-
Data is sent incomplete, to the wrong device, or not at all.
-
Patient data is blocked, falsified, or corrupted.
-
The time stamp of the data is inaccurate.
Overloading this unit due to very high network loading (e.g. denial of service attacks) can
cause interface deactivation. The interface can only be used again after the device is
restarted. Rarely, booting may be slow or repeated reboots may occur.
Содержание BM1
Страница 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Страница 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Страница 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Страница 55: ...Rev 2 2 55 ...
Страница 62: ...Rev 2 2 62 Alarm Review ...
Страница 66: ...Rev 2 2 66 Click the numeric window you want to edit Edit total score items Edit numeric items Edit text item ...
Страница 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Страница 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...