Rev. 2.2
93
the patient's blood pressure value.
Every Time: The set pressure is applied
every time blood pressure is measured.
Whenever blood pressure is measured,
pressurize to the set pressure value.
A-4.
Auto M
EASUREMENT
INTERVAL
A menu to set Interval time when
measures the blood pressure
periodically.
Interval range : 1min, 2, 3, 4, 5, 10, 15,
20, 30, 1hour, 2, 4, 8.
After setting INTERVAL, you must press
NIBP KEY to start NIBP START
periodically.
1min, 2, 3, 4, 5, 10,
15, 20, 30, 1hour, 2,
4, 8
B-1.
NIBP STAT
Patients with severe state changes in
blood pressure are in continuous
mode for 5minutes to check for
changes in blood pressure
continuously.
OFF / ON
C-1.
VITAL SIGN REVIEW
Record the 40 most recently measured
blood pressure values.
NIBP Status Messages
Messages
Status
OVER PRESSURE
When the cuff pressure is excessive
OVER TIME CUFF PRESSURE
When the cuff pressure exceeds the set
time
INFLATION FAILURE CHECK CUFF
If the cuff hose is not connected properly
DEFLATION FAILURE
When the cuff breaks and cannot exhaust
MEASUREMENT ERROR
When there is no measurement signal
Содержание BM1
Страница 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Страница 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Страница 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Страница 55: ...Rev 2 2 55 ...
Страница 62: ...Rev 2 2 62 Alarm Review ...
Страница 66: ...Rev 2 2 66 Click the numeric window you want to edit Edit total score items Edit numeric items Edit text item ...
Страница 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Страница 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...