Rev. 2.2
109
type is changed.
MODULE OFF
It occurs when the equipment and
module are separated. Message
output
Verify module connections
, Service request
EtCO2 Measurement Failure
CO2 value is not output, or numerical error. Troubleshoot procedure
1.
Check the connection between the main unit and the module
2.
Check the module line connection with the filter line or airway
3.
Replace filter line or airway
4.
Service Request
NOTE
In the following monitoring conditions, the measured values may be inaccurate. Read the
measured values carefully.
1.
When using this in an environment of using Nitrous Oxide gas of high concentration
2.
When using this in an environment where abrupt temperature change takes place
3.
When using this in an environment with severely high humidity
CAUTION
The measured values may be inaccurate when using this equipment for patients who
have very fast or irregular respiration.
When measuring CO2 from the patient under the anesthesia, check it when gas
mixture comes in. Otherwise, the measured result values may be inaccurate.
When using a anesthesia machine that uses a volatile anesthetic, CO2 values may be
inaccurate.
Содержание BM1
Страница 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Страница 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Страница 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Страница 55: ...Rev 2 2 55 ...
Страница 62: ...Rev 2 2 62 Alarm Review ...
Страница 66: ...Rev 2 2 66 Click the numeric window you want to edit Edit total score items Edit numeric items Edit text item ...
Страница 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Страница 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...