Rev. 2.2
92
NIBP Settings
Menu
Description
Available Settings
A. NIBP
A-1.
ALARM
NIBP Alarm setup menu
A-1-1.
PARAMETER ALARM
LIMIT
SYS, MEAN, DIA, PR Parameter alarm
limit, level and activation setup
A-1-2.
TECHNICAL ALARM
CONDITION
NIBP-OVER PRESSURE
NIBP-OVERTIME PRESSURE
NIBP-INFLATION FAILURE
NIBP-DEFLATION FAILUER
NIBP-MEASUREMENT ERROR
NIBP-PULSE TOO WEAK
NIBP-AIR LEAK
NIBP-EXCESSIVE MOTION
NIBP-SYSTEM FAULT
A-2.
INFLATION
It is a function to set the range that is
usually used by setting pressure at the
beginning because it can give pain to
the patient when the equipment is
turned on and pressurized to the
maximum pressure range at the initial
pressurization.
Default Settings value:
ADT: 170 mmHg
PED: 140mmHg
NEO: 120mmHg
ADT : 120 – 250
mmHg
PED : 80 – 170mmHg
NEO : 60 –
140mmHg
A-3.
Setting Time
How to apply pressurization settings.
Once: Applies the set pressure value
only once when booting the
instrument and registering a patient
When the blood pressure is measured
for the first time, the pressure is set to
the set pressure value, but
automatically adjusted according to
Once, Every Time
Содержание BM1
Страница 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Страница 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Страница 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Страница 55: ...Rev 2 2 55 ...
Страница 62: ...Rev 2 2 62 Alarm Review ...
Страница 66: ...Rev 2 2 66 Click the numeric window you want to edit Edit total score items Edit numeric items Edit text item ...
Страница 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Страница 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...