Rev. 2.2
86
With obesity, where a thick layer of fat surrounding a limb dampens the oscillations
coming from the artery
On an edematous extremity
The effectiveness of this sphygmomanometer has not been established in pregnant,
including preeclamptic patients
WARNING
Non-invasive blood pressure monitoring is not recommended for patients with hypotension,
hypertension, arrhythmias or extremely high or low heart rate. The software algorithm cannot
accurately measure NIBP for patients with these conditions.
WARNING
As the value of NIBP can vary according to the age and sex of a patient, the user needs to
set up right data in parameter Menu before measurement.
Tubes between the cuff and the monitor are not kinked or blocked.
Pay attention to not to block connecting hose when you put cuff on patient.
Cuff or hose connection for leaks periodically. Measurements can be inaccurate if air leaks.
The air pad should be exactly over the branchial artery. Tubing is immediately to the right
or left of the branchial artery to prevent kinking when elbow is bent.
Try to measure infants when they are calm. A kicking or crying baby may disturb or jiggle
the cuff, causing noise within the system and resulting in unstable blood pressure readings.
If necessary, hold the cuffed limb steady, without impeding circulation. Do not hold onto
the cuff and do not pat the cuffed limb to comfort the child.
NIBP cannot be taken under all conditions. Even manual methods, employing a
sphygmomanometer and stethoscope, will not work on unstable or active patients.
Содержание BM1
Страница 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Страница 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Страница 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Страница 55: ...Rev 2 2 55 ...
Страница 62: ...Rev 2 2 62 Alarm Review ...
Страница 66: ...Rev 2 2 66 Click the numeric window you want to edit Edit total score items Edit numeric items Edit text item ...
Страница 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Страница 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...