12. Maintenance and
Troubleshooting
Inspection Equipment
You should perform a visual inspection before every use, and in accordance with your hospital’s
policy. With the monitor switched off:
Examine unit exteriors for cleanliness and general physical condition. Make sure that the
housings are not cracked or broken, that everything is present, that there are no spilled
liquids and that there are no signs of abuse.
If the EtCO2 module is mounted on the monitor, make sure that it is locked into place and
do not slide out without releasing the locking mechanism.
Inspect all accessories (cables, transducers, sensors and so forth). If any show signs of
damage, do not use.
Switch the monitor on and make sure the backlight is bright enough. Check that screen is at its full
brightness. If the brightness is not adequate, contact your service personnel or your supplier.
Inspection Cables
Examine all system cables, the power plug for damage. Make sure that the prongs of the
plug do not move in the adaptor. If damaged, replace it with an appropriate Bionet power
cord and adaptor.
Inspect the parameter cable and ensure that it makes good connection with the Monitor.
Make sure that there are no breaks in the insulation.
Apply the transducer or electrodes to the patient, and with the monitor switched on, flex
the Patient cables near each end to make sure that there are no intermittent faults
Содержание BM1
Страница 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Страница 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Страница 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Страница 55: ...Rev 2 2 55 ...
Страница 62: ...Rev 2 2 62 Alarm Review ...
Страница 66: ...Rev 2 2 66 Click the numeric window you want to edit Edit total score items Edit numeric items Edit text item ...
Страница 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Страница 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...