Rev. 2.2
104
When you select a range value, the
selected pressure range value is
displayed below the dotted line
above the two dotted lines in the left
middle of the WAVE window.
100mmHg (13.2vol%)
150mmHg (19.7vol%)
B. EtCO2 Menu In Text Window
Menu
Description
Available settings
B. EtCO2 Parameters
B-1.
ALARM
EtCO2 Alarm Setup Menu
B-1-1.
PARAMETER ALARM
LIMIT
ETCO2, FICO2, AWRR, APNEA
parameter alarm, level, action setup
menu
B-1-1-1.
ETCO2-ETCO2
Alarm
Level
Upper/Lower
On/Off
Message/Low/Medium
/High
0~100(mmHg)
B-1-1-2.
ETCO2-FICO2
Alarm
Level
Upper/Lower
On/Off
Message/Low/Medium
/High
0~20(mmHg)
B-1-1-3.
ETCO2-AWRR
Alarm
Level
Upper/Lower
On/Off
Message/Low/Medium
/High
0~150(RPM)
B-1-1-4.
ETCO2-APNEA
Alarm
Level
Upper
On/Off
Message/Low/Medium
/High
10~60(s)
B-1-2.
TECHNICAL ALARM
CONDITION
ETCO2-MODULE OFF
ETCO2-CHECK ADAPTOR
ETCO2-CHECK LINE
ETCO2-CHEKC LINE DISCONNECT
ETCO2-CO2 INVALID
ETCO2-OVER RANGE
ETCO2-ZERO REQUIRED
Alarm: On/Off
Level:
Message/Low/Medium
/High
Содержание BM1
Страница 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Страница 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Страница 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Страница 55: ...Rev 2 2 55 ...
Страница 62: ...Rev 2 2 62 Alarm Review ...
Страница 66: ...Rev 2 2 66 Click the numeric window you want to edit Edit total score items Edit numeric items Edit text item ...
Страница 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Страница 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...