Rev. 2.2
132
SpO2 Performance
Saturation range
0 to 100%
Saturation accuracy
70 to 100% ±2 digits
0 to 69% unspecified
Pulse rate range
30 to 254 bpm
Pulse rate accuracy
±2 bpm
NIBP Performance
Method
Oscillometric with linear deflation
Operation Mode
Manual/Automatic/Continuous
Measurement range
Adult Pressure : 20 to 260 mmHg
Pediatric Pressure : 20 to 230 mmHg
Neonate Pressure : 20 to 120 mmHg
Accuracy
mean error: less than ±5 mmHg
standard deviation : less than 8 mmHg
Meets accuracy requirements of ANSI/AAMI SP10:1992 and 2002
Temperature Performance
Measurement range
34 to 42.5℃ (93.2 to 108.5℉)
Accuracy
36℃to 39℃: ±0.2℃
Other: ±0.3℃
Sidestream CO2 (Option)
Measurement range
0 to 150 mmHg, 0 to 19%
Accuracy
0-40mmHg ±2 mmHg,
41-70mmHg ±5% of reading
71-100mmHg ±8% of reading,