Rev. 2.2
107
balance gas is set to helium.
B-5-8.
ZERO TYPE
When performing a zero on room air,
this setting should be set to room air
(the default). Only change to nitrogen
(N
2
) when performing a zero on
100% N
2
gas; this is provided for use
in a laboratory environment.
ROOM AIR / N2
C-1.
ZEROING
This function is used to initiate a
Capnostat Zero.
A zero is used to correct for
differences in airway adapter types.
The Capnostat zero must be
performed free of any CO2
1.
Set the Host to the zeroing
function.
2.
Connect the CAPNOSTAT 5 CO2
Sensor
3.
Place the CAPNOSTAT 5 CO2
Sensor onto a clean and dry CO2
adapter that is exposed to room
air and away from all sources of
CO2, including the ventilator, the
patient’s breath and your own.
Start the adapter zero. The maximum
time for a CAPNOSTAT zero is
40seconds. The typical time for a
zero is 15~20seconds.
D-1.
MODULE RESET
EtCO2 MODULE initializing.
NOTE
For best result, connect the CAPNOSTAT 5 CO2 Sensor to an adapter and wait 2minutes before
performing the Adapter Zero procedure.
Содержание BM1
Страница 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Страница 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Страница 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Страница 55: ...Rev 2 2 55 ...
Страница 62: ...Rev 2 2 62 Alarm Review ...
Страница 66: ...Rev 2 2 66 Click the numeric window you want to edit Edit total score items Edit numeric items Edit text item ...
Страница 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Страница 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...