Connecting this device to an integrated network with other devices, or subsequent changes to that
network, can be a new risk to patients, users, and third parties. These risks must be identified,
analyzed and evaluated before the device is connected to the network or the network is changed,
and appropriate action must be taken.
Subsequent changes to the network example:
Network configuration change
Removing a device from the network
Adding new devices to the network
Upgrading or updating devices connected to the network
WARNING
Recommendations for wireless connections
BM1 has a change in the number of equipment connections depending on wireless
AP (Access Point) performance.
When using a general AP, it is recommended to connect 8 units to the same network.
It is recommended to use the AP exclusively for monitoring equipment.
Due to the nature of wireless, connectivity may not be good depending on the
environment.