Rev. 2.2
128
monitor.
Portable and mobile RF communication equipment can affect medical electrical
equipment.
Cables and accessories not specified in the user guide are not certified. Using other
cables and / or accessories may adversely affect safety, performance, and
electromagnetic compatibility (increased electromagnetic emissions and reduced
immunity).
This equipment should not be used near or on top of other equipment. If you need
to use it on its side or stacked, you should observe the equipment to make sure it
works properly within your configuration.
This patient monitoring device communicates over a 2.4 GHz 802.11b / g wireless
network. Other equipment may interfere with data reception on this wireless network.
This is also true if the equipment complies with the CISPR emission requirements.
When using patient monitoring equipment to communicate over a wireless network,
be sure to check that it is compatible with existing or new wireless systems (eg, cell
phones, pager systems, cordless phones, etc.). For example, a Bluetooth-compliant
device using the 2.4 GHz frequency band may interfere with the wireless
communication of the patient monitor. For more information on wireless deployment,
please contact your Bionet representative.
Low amplitude signals such as EEG and ECG are particularly sensitive to interference
from electromagnetic energy. This equipment complies with the tests listed at the
bottom, but does not guarantee complete operation. The "quiet" electrical
environment is better. In general, the greater the distance between electrical
equipment, the lower the likelihood of interference.
Содержание BM1
Страница 14: ...Rev 2 2 14 CAUTION Infectious devices and parts must be sanitized and cleaned before disposal ...
Страница 19: ...Rev 2 2 19 Basic Unit Front View Alarm lamp Battery operation indicator AC status indicator ...
Страница 22: ...Rev 2 2 22 Back Side View Speaker Hole IV Pole Mount Battery Cover ...
Страница 55: ...Rev 2 2 55 ...
Страница 62: ...Rev 2 2 62 Alarm Review ...
Страница 66: ...Rev 2 2 66 Click the numeric window you want to edit Edit total score items Edit numeric items Edit text item ...
Страница 89: ...Rev 2 2 89 NIBP Connector and Measurement Cable NIBP Connector Adult Cuff ...
Страница 139: ...Rev 2 2 139 Symbols and degree s greater than less than minus number percent plus or minus ...