Chapter 7 Commissioning the Ikus and setting parameters
INSTRUCTION
1.
Remove the seal plugs from the driving tube connection socket.
2.
Connect the plug of the 2 tank units to the driving tube connection sockets.
7.1.2 Switching on the Ikus
WARNING
The
Ikus
must always be connected to the power supply when it is
switched on. This is the only way to ensure that the start-up test (see
section 7.1.3: Starting the monitor program, page 95) is performed
completely and possible malfunctions can be detected.
CAUTION
When switching on always take care to switch on the
Ikus
first and then
the laptop. Never vice versa!
INSTRUCTION
1.
Connect the Ikus to the mains. Secure the mains cable with the plug clip.
Ensure that the mains power switch (toggle switch) is set to
[I]
position.
2.
Turn the main switch (key switch) to the
[I]
position. The battery charge indica
tor will light up and the number of hours the driving unit has been operated will
be displayed. The mains operation indicator lights up.
3.
The menu
Select language
appears after the laptop is switched on.
4.
Select the desired language by pressing the corresponding number key. It is
not necessary to press
<Enter>
to confirm this selection.
5.
The start menu is displayed on the laptop.
7.1.3 Starting the monitor program
INSTRUCTION
1.
Select the
1. Start program
option (
<1>
) in the start menu.
2.
Enter user ID and password, confirm with
<Enter>
. The
Ikus
will carry out a
start-test.
3.
Wait for the start-test to finish (it takes several minutes). Do not mute the
acoustic signal by pressing the
Mute button
. Exception: an error message is
displayed. The messages in the message window will provide information on
the status of the test. If the
Ikus
is found to be operating correctly, the view
Select operating mode
will be displayed next.
7.1.4 Setting the test parameters
NOTICE
When delivered, the Ikus has standard default parameters that it uses
after each complete start-up process (see section 6.5.1: Routine start-
test when not in operation, page 83 and section 15.3.2: Ikus start-test
following emergency pulse mode, page 201). However, during the
warm-up period with the tank unit, it is necessary to set the test
parameters to prevent the Ikus from generating a false error message.
1000721x09 Revision 8
95
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...