Chapter 1 Important safety information
Only disconnect the Ikus from the mains if the charge level indicator
shows that the batteries are completely charged (all yellow LEDs are
lit).
IMPORTANT: In order to prevent rapid and premature ageing of the
batteries, the Ikus must always be run on the mains for at least 6 hours
after using the battery operation. Only after this may the Ikus be run on
battery operation again.
If the LEDs of the charge level indicator are blinking or the message
Batteries discharged - use power supply!
appears, immediately
switch to mains operation! If the batteries are completely discharged
(red LED of the charge level indicator is lit) and the drive continues to
run on battery operation, there is the risk of a total malfunction of the
Ikus and damage of the batteries. If this happens, it can not be
guaranteed that the Ikus will restart after connecting it to the mains.
Whenever the Ikus is running in battery operation, the patient must be
accompanied by a person trained to use the manual pump. Thus the
patient shall be guaranteed care in an emergency.
To prevent the batteries from aging quickly, every period of battery
operation should be followed by at least 6 h of mains operation.
When the battery charge is low, the acoustic signal sounds every
minute. The Ikus must be connected to the mains operation
immediately.
Messages are only displayed when the monitor program is running.
When the monitor program is shut down, the only indications that there
is an error message are an acoustic signal and the fact that the
indicator lamp on the Ikus handle lights up. There is no way of finding
out what type of error message has been displayed.
When reading out the log files: Make sure that you always have the
USB stick inserted and that there is sufficient capacity on the stick.
Otherwise the LOG files might get lost as they are deleted from the hard
disk as soon as they have been transferred onto a USB stick.
1.1.5 Packaging and sterilization
WARNING
EXCOR blood pumps and cannulae are intended for single-use only.
Otherwise there is a risk of infection.
1000721x09 Revision 8
7
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...