Chapter 14 Error Messages and corrective measures
HOTLINE
Notify Berlin Heart! 866.249.0128
14.6.2 Emergency operation system 1 (or system 2 or 3). Contact Service
now!
•
Emergency operation System 1. Contact service now!
•
Emergency operation System 2. Contact service now!
•
Emergency operation System 3. Contact service now!
WARNING
There is no longer a redundant backup system. If the only remaining
intact system fails, there is a risk that the Ikus will stop running
altogether.
Appears in biventricular mode
Both blood pumps are being driven by the last intact pneumatic system.
The pneumatic system will now operate with fixed parameters (synchronous mode,
systolic pressure 250 mmHg, diastolic pressure -100 mmHg, 70 bpm, relative systolic
duration 40 %). It is not possible to change these settings.
Possible causes:
•
2 of the 3 pneumatic systems have developed faults.
•
The Ikus driving unit has been running on battery power for too long (Error
message:
Emergency operation due to empty batteries: Risk of total
failure!
). It was no longer possible to establish reliable conditions. In order to
ensure that this is not due to a control circuit defect, the Ikus has switched over
to the backup system and the backup control computer. The message
Backup
computer started! Contact customer service.
is displayed.
INSTRUCTION
1.
Provide the patient immediately with a replacement Ikus.
HOTLINE
Notify Berlin Heart! 866.249.0128
14.7 System 1 (or system 2) is defective!
•
System 1 (left) is defective!
•
System 2 (right) is defective!
The respective pneumatic system has developed a fault. The backup system (system
3) is activated. If this error occurs during backup operation, Ikus will be running in
emergency operating mode. In this case, Ikus has no more redundancy.
INSTRUCTION
1.
Provide the patient immediately with a replacement Ikus.
HOTLINE
Notify Berlin Heart! 866.249.0128
1000721x09 Revision 8
177
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...