Chapter 14 Error Messages and corrective measures
10.
The system starts up again using the defined parameters.
11.
Check whether the pump is filling correctly and, if necessary, adjust the Ikus
parameters.
12.
Secure the pump end of the driving tube with a cable tie strap. Important: Only
the cable ties and cable tie guns provided should be used. See section 9.11:
Securing the connections, page 130.
14.4 Please check left / right pump and driving tube
•
Please check left pump and driving tube!
•
Please check right pump and driving tube!
The Ikus has detected an excessively deviating flow: The mean values of the last
16 cycles and of the last 1024 cycles deviate by more than 10 % from each other.
Depending on the components, the parameters and the cause of the message, the
message will be displayed usually within 5 s to 1 min.
WARNING
Rates < 60 bpm are intended to be used only for implantation and
explantation. Never use the Ikus with a rate < 60 bpm without constant
supervision.
IMPORTANT: First mute the message on the drive unit, do not acknowledge yet in the
monitor program. Otherwise, alarms will occur by mistake. When muted, the alarm will
be audible again in an 1-minute interval.
INSTRUCTION
1.
Mute the message on the drive unit.
2.
Inspect the driving tube and the cannulae: are kinks blocking the flow? Correct
the positions to ensure an unimpaired flow.
3.
Inspect the driving tube and the plugs. If a plug is not positioned correctly: re
insert it correctly. If the driving tube is defective: replace it (see section 14.3.1:
Replacing a driving tube, page 174).
4.
If necessary, adjust the parameters. Wait until the message
Left: Pump out
put measurement activated
or
Right: Pump output measurement acti
vated
appears.
5.
If necessary, correct the position of the cannulae.
6.
Assess the hemodynamic status of the patient (volumes, MAP, PAP, CVP,
etc.).
If the message appears again:
INSTRUCTION
1.
Mute the message on the drive.
2.
Confirm the parameter values. For that the cursor must be in the parameter
table.
3.
Assess the patient’s hemodynamic status (volume, MAP, PAP, CVP...)
1000721x09 Revision 8
175
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...