Chapter 14 Error Messages and corrective measures
14.3 Please connect driving tube
Please connect driving tube!
INSTRUCTION
1.
Inspect the driving tube and the connectors.
2.
If a plug is not seated correctly: re-insert it correctly. To do so, grip the release
sleeve and pull it out of the connector, then plug it back in. Check that the plug
is securely connected. To do so, grip the plug body above the release sleeve
and pull on it. Do not pull from the release sleeve, and never from the tube!
3.
If the driving tube is defective: replace it.
14.3.1 Replacing a driving tube
The driving tubes (red/blue) have a maximum life of 1 year (see section 16.2:
Technical specifications, page 217) and must be replaced after 1 year.
WARNING
In order for a driving tube to be replaced, the pump must be stopped for
a short time. If the left driving tube is being replaced in a driving unit
providing biventricular support, the right pump must also be stopped
while the driving tube is being replaced in order to avoid overloading of
the pulmonary circulation (danger of pulmonary edema).
Material
• 1 driving tube, red or blue
• 1 tube connecting set (cable tie, cable-tie gun), from accessory set
INSTRUCTION
1.
If required log into the monitor program by entering user ID and password and
confirm with
<Enter>
.
2.
Select the option
Drive pause
and press
<Enter>
to confirm. Respond to the
prompt in the dialog window by pressing the
<X>
key or the
<1>
key. The Ikus
will stop.
3.
Carefully cut the cable tie on the defective driving tube.
4.
As soon as the pump has stopped, remove the defective driving tube from the
pump.
5.
Connect the new driving tube to the blood pump. To do so, carefully push the
end of the driving tube onto the driving tube connector.
6.
Remove the defective driving tube from the Ikus connector socket. To do so,
take hold of the release sleeve and pull this out of the connector.
7.
Connect the new driving tube to the connector, which is now free. The sound
of the plug snapping into place is clearly audible. Check that the plug is
securely connected. To do so, grip the plug body above the release sleeve and
pull on it. Do not pull from the release sleeve, and never from the tube!
8.
The view
Select operating mode
appears on the monitor. Select
Univentricu
lar (UVAD)
or
Biventricular (BVAD)
, then confirm the selected operating
mode with
<Enter>
. In biventricular mode the view
Pump size and single-step
mode
appears. In univentricular mode, the connector seal test is performed
first and then the view
Pump size and single-step mode
appears.
9.
To confirm the
OK
selection, press
<Enter>
.
1000721x09 Revision 8
174
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...