Chapter 11 Wound care and treatment
Adjusting the parameter values
Only adjust the parameters if the measures listed above have no effect or in case of:
•
Mobilization of patient: adjust the systolic pressure, both left and right. When
pressures have increased, do not reduce these again, even when the patient is
lying down.
•
Signs of low cardiac output: the membrane is moving properly while at the same
time a decrease in urine output, lactate increase and dyspnea (shortage of
breath) can be observed. In this case, increase the rate and adjust other settings
as required.
INSTRUCTION
1.
Use the
<
←
>
/
<
→
>
keys to move the cursor to the desired field in the parame
ter table. The selected field is given a colored background.
2.
Use the
<
↓
>
,
<
↑
>
and
<Bild-
↓
>
,
<Bild-
↑
>
keys to adjust the value, then press
<Enter>
to confirm the input. The system will now operate using the new set
tings.
Cautionary measure
Confirm each changed parameter value by pressing
<Enter>
. The system does not
take over the new, changed value until it has been confirmed with
<Enter>
.
ADVICE
Enter all the changes to the parameter values into the parameter log.
(see section 16.7: Pump performance flow sheet, page 230).
11.5.2 Visual inspection: deposits
Check the blood pump(s) and the visible part of the cannulae (cannula extension set
/ connecting set resp.) for visible deposits (fibrin, clots) every 4 hours. If deposits
develop, check the pump(s) every hour.
1000721x09 Revision 8
142
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...