Chapter 1 Important safety information
Only connect the mains cable of the Ikus driving unit to grounded mains
power outlets.The supply voltage has to conform to the voltage
requirements indicated on the Ikus identification plate. Only connect
the mains cable of the Ikus driving unit to suitable power outlets
complying with the electrical safety regulations of the country in which
it is being used. Otherwise there is a risk of electrical shock or damage
to the EXCOR system!
If there are several electrical devices in the vicinity of the patient, then
these are to be connected to a central grounding point. The connector
Potential equalization
is to be used for this on the Ikus. Otherwise there
is a risk of electric shock.
Switch on the Ikus 2 hours before use in order to sufficiently charge the
batteries and to detect possible device errors during the start test see
chapter 7.1.2: Switching on the Ikus, page 95.IMPORTANT: During
this time, always connect both tank units to the Ikus (see section 7.1.1:
Connecting the tank unit, page 94)! Otherwise there is a risk that error
messages are falsely generated.
After switching on the Ikus, pull the key out of the main switch (key
switch) and store it in a safe place.
The Ikus power switch (toggle switch) must be switched on during the
first startup and remain set to
[I]
position. Its position should never be
changed. Otherwise there is a risk that the batteries will not be fully
charged after battery operation.
The Ikus is designed for stationary operation and to be run on mains
power (referred to as the mains in this instruction for use). Do not run it
on battery operation unless this is absolutely necessary (e. g. when
moving the patient within the clinic or during a mains failure).
Always take the battery capacity limits into consideration when using
the system. When the system is working in battery operation, the
patient must be accompanied by a person trained to use the manual
pump. Thus the patient shall be guaranteed care in an emergency.
The air vents must not be covered or obstructed during operation.
Otherwise the Ikus will not receive sufficient ventilation, which may lead
to overheating. This may lead to malfunctions and a malfunction in
device operation.
1000721x09 Revision 8
6
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...