Chapter 13 Weaning and Explantation for BTR and BTT
Parameter
Explanation
Abbr.
Value
total weaning
interval
Difference between
initial rate and
explantation rate:
TWI = IR - WR
TWI
Please enter:
IR ___ bpm - WR 30 bpm =
TWI ___ bpm
reduced rate
rate resumed at the
end of day 1 to 3
RR
1
to
RR
3
Please refer to Tab. 13-8,
page 163.
Tab. 13-7
Important parameters for weaning progress
Reduced
rate (RR
x
)
Calculation
RR
1
Please enter:
RR
1
= WR 30 bpm + 0.75 x TWI ( ___ bpm) = ____bpm
RR
2
Please enter:
RR
2
= WR 30 bpm + 0.50 x TWI ( ___ bpm) = ____bpm
RR
3
Please enter:
RR
3
= WR 30 bpm + 0.25 x TWI ( ___ bpm) = ____bpm
Tab. 13-8
Reduced rate day 1 to day 3
50 / 60 ml pump Weaning Sequence
50 / 60 ml pump Weaning Sequence
Day 0
After confirmation of eligibility criteria, the following steps should be
performed under echo guidance:
1
1.
Administer unfractionated heparin (UFH) 75 units/kg x _____ kg =
_____ mg IV x 1 [max 5000 units].
2.
After 5 minutes, reduce the pump rate step-wise from IR (_____ bpm)
to 30 bpm in increments of 5 bpm q5 min. After 5 minutes at 30 bpm,
reassess LV size and function.
3.
After an additional 5 minutes (i.e. total time = 10 min at 30 bpm), stop
the pump for 5 min and reassess LV size and function. During pump
stoppage, use the manual pump to pump twice q30 seconds to
minimize the risk of thrombus formation due to blood stagnation while
Ikus is disconnected.
4.
After 5-minute pump stop, reconnect pump to Ikus and resume pump
speed at IR(___ bpm).
Did the patient satisfy all 5 eligibility criteria throughout
the period of weaning and pump stoppage?
ÿ
NO - STOP
ÿ
YES -
Proceed
MD________
Tab. 13-9
50 / 60 ml pump Weaning Sequence
1000721x09 Revision 8
163
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...