Chapter 7 Commissioning the Ikus and setting parameters
INSTRUCTION
1.
When the blood pump(s) starts to fill, reduce the CPB flow and gradually
increase the EXCOR rate from an initial 30 bpm until CPB has been termi
nated and the required flow is achieved. IMPORTANT: In doing so, make sure
that the pump fills adequately, and if necessary regulate the driving pressure.
2.
If necessary, adjust the systolic pressure, diastolic pressure and the systolic
percent.
7.2.12 Possible complications
Decreased filling after stable filling conditions
If a good filling behavior was achieved at first (filling pressures LA/CVP < 10 mmHg
and diastolic pulmonary artery pressure < 15 mmHg) with good drainage and nominal
rate (normally 80 bpm), but the filling has deteriorated over time, it usually will not help
to increase the diastolic pressure.
Deterioration in the filling behavior despite stable inflow conditions may indicate
hypovolemia or obstruction of the inflow cannula. The cause of deterioration in filling
behavior must be identified and addressed.
NOTICE
Manipulations during implantation can severely influence the inflow
temporarily – wait for the situation to stabilize before adjusting the
values.
INSTRUCTION
1.
Evaluate volume status and transfuse if necessary. Evaluate and if necessary
correct the cannula position.
Pump filling deteriorates when thorax is closed
If atrial cannulation is used, a slight decrease in the filling may be observed in some
cases when the thorax is closed. This may be caused by compression of the atria or
a slight shift in the position of the cannulae.
INSTRUCTION
1.
Evaluate volume status and transfuse if necessary. IMPORTANT: Observe the
effect volume replacement on the pump filling!
2.
Increase suction pressure.
Distinct decrease in filling or generally poor inflow conditions on right side
INSTRUCTION
1.
Make sure that there is no inflow obstruction.
2.
If a suction pressure of less than -50 mmHg is necessary, increase the relative
diastolic duration as an additional measure. At the same time, the systolic
duration is reduced accordingly. IMPORTANT: Increase the driving pressure
accordingly!
1000721x09 Revision 8
106
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...