Chapter 1 Important safety information
1
Important safety information
1.1
Warnings
WARNING
Before using EXCOR, read the Instructions for Use carefully.
Only qualified medical personnel trained specifically in the use of the
system are permitted to work with EXCOR. Training courses can be
arranged with Berlin Heart, Inc. Use by untrained personnel can pose
a risk to the patient and the EXCOR.
Before starting the Ikus, make sure that a replacement Ikus is available
in the hospital. If a replacement Ikus is not available, there is a risk that
the patient cannot be cared for in the event of device malfunction.
The general rule is:
1 replacement Ikus if 1 or 2 systems are in use,
2 replacement Ikus if 3 or 4 systems are in use,
3 replacement Ikus if 5 or 6 systems are in use.
If more than 6 systems are in use the number of replacement Ikus has
to be 1/2 of the active systems.
On the system EXCOR only use components of this system. Never use
other components than those delivered by Berlin Heart GmbH/ Berlin
Heart Inc. Otherwise the warranty is no longer valid.
The Ikus and the components of the EXCOR system must not be
modified. Otherwise the secure function of the system can not be
guaranteed.
The Ikus should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the Ikus should be
observed to verify normal operation in the configuration in which it will
be used.
The system EXCOR Pediatric and its components are permitted to be
used only by prescription of the attending physician.
To ensure the safety of the patient supported with the EXCOR pediatric
the patient should be supervised by qualified medical personnel who
have been trained on the use of the system.
Do not use the EXCOR if there is any visable damage of the Ikus or any
of its components.
If there is any malfunction of the Ikus while the driving unit is connected
to the patient, the Ikus must immediately be replaced.
1000721x09 Revision 8
3
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...