Chapter 6 Instructions for use: Ikus
Instructions for use: Ikus
WARNING
Do not install any other software on this laptop.
Make sure that the
<NumLk>
and the
<Caps Lock>
key on the laptop
is deactivated. The status LED on the laptop, identified by a lock
symbol and/or a numeral (e. g. 1) should be off (see Fig. 6-1, page 67
to Fig. 6-3, page 68)
Only use the USB stick provided with the device for saving data. Never
connect any other USB stick to the laptop!
The laptop must be switched off before the USB stick is inserted in or
removed from the USB port.
Check all the information and messages in the message window of the
monitor program at least every 4 hours. If there is a message that
indicates action must be taken, take the appropriate action and, if
necessary, contact the Berlin Heart, Inc. emergency hotline.
When switching on the Ikus, always switch the Ikus on first and then the
laptop and not vice versa.
NOTICE
For details on commissioning the system see chapter 7:
Commissioning the Ikus and setting parameters, page 93.
The Ikus is controlled and monitored using the monitor program, which is permanently
installed on the laptop. Only users who are properly logged in can use the monitor
program to change system settings.
Fig. 6-1
LEDs below the mouse pad
1
LED for
<Caps Lock>
2
LED for
<NumLk>
1000721x09 Revision 8
67
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...