Chapter 15 Troubleshooting and correcting faults
Systole
[mmHg]
left/ right
Diastole
[mmHg]
left/ right
Rate
[bpm]
Rel. diast.
duration [%]
left/ right
Operation mode
210/ 130
-40/ -20
80
40/40
biventricular,
synchronous and
separate
Tab. 15-3
Default standard parameters
INSTRUCTION
1.
Select the desired language by pressing the corresponding number key. It is
not necessary to press
<Enter>
to confirm this selection. The start menu is
displayed on the laptop.
2.
Select the
1. Start program
option (
<1>
). Enter user ID and password, con
firming the password with
<Enter>
. The
Ikus
will carry out a start-test.
3.
Wait for the start-test phase to finish (this takes a few minutes). Do not mute
the acoustic signal. The messages in the message window inform the user of
the current status of the test. If the driving unit is found to be operating cor
rectly, the view
Select operating mode
will be displayed next.
4.
Select
Univentricular (UVAD)
or
Biventricular (BVAD)
, then confirm the
selected operating mode with
<Enter>
.
5.
In
biventricular
mode, the view
Pump size and single-step mode
is shown. In
univentricular
mode, a connector seal test is first performed (taking approx.
10 seconds). The
Ikus
checks whether the driving tube connector with the
blue marking has been sealed. Then the view
Pump size and single-step
mode
is displayed.
6.
Transfer all the parameters from the original
Ikus
to the replacement
Ikus
.
7.
Move cursor to the field
Step left
and press
<Enter>
to confirm.
Biventricular
: Move cursor to the field
Step right
and press
<Enter>
to con
firm.
8.
Move the cursor to the
OK
field. Important: Do not confirm
OK
yet.
The default settings for systole, diastole and relative systole duration depend on
whether the pump is registered as the left or right pump in the monitor program.
Univentricular
: Default setting as for left pump.
Connecting the blood pump(s) to the replacement Ikus
INSTRUCTION
1.
For
univentricular operation
, unplug the connector marked in red, for
biven
tricular operation
, unplug both connectors. To do so, remove the seal plugs
from the respective connector(s).
2.
If possible
, log into the monitor program of the original
Ikus
. Select
Drive
OFF
, then press
<Enter>
to confirm. Respond to the prompt in the dialog win
dow by pressing the
<X>
key or the
<1>
key. The system stops immediately. If
it is not possible to select the
Drive OFF
option, proceed with triggering single
steps (
Step left
/
Step right
, see instructions in section 15.1: Replacing the
blood pump(s), page 195) without stopping the
Ikus
.
1000721x09 Revision 8
203
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...