Chapter 15 Troubleshooting and correcting faults
Problem
Cause of problem / action to be taken
Cannula rupture
Possible cause:
Damage to cannula caused by excessive external forces
or a sharp object.
What to do?
Immediately stop the support by disconnecting the driving
tube from the Ikus.
Clamp the cannula.
Follow directions to replace the pump, see section 15.1:
Replacing the blood pump(s), page 195.
After removing the current pump, trim the cannula
proximal to the damaged area.
Defective driving
Possible causes:
tube(s) found
through visible or
audible inspection
of the tube(s) or
from Ikus alarm(s)
The driving tube has been damaged and the integrity of
the tubing may be compromised. There may or may not
be an audible sound from air escaping the tubing. This
condition may be accompanied by the following error
messages.
• section 14.1: Pressure error / time error in system 1
(or in system 2 or 3), page 173 followed by:
• section 14.3: Please connect driving tube, page 174
What to do?
If the driving tube is defective, replace it; see
section 14.3.1: Replacing a driving tube, page 174.
If a fault in the tubing is not apparent upon inspection, but
the Ikus has the above error(s), then follow the directions
associated with the error message on the Ikus; see
chapter 14: Error Messages and corrective measures,
page 171:
Visible Ikus faults
Notify Berlin Heart.
Tab. 15-1
Possible problems
1000721x09 Revision 8
193
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...