Chapter 14 Error Messages and corrective measures
WARNING
Danger of total malfunction of the Ikus in the event of an additional
error! Take immediate action!
INSTRUCTION
1.
Provide the patient immediately with a replacement Ikus and switch off the
malfunctioning Ikus. (see section 15.4: Connecting the patient to a replace
ment Ikus, page 202). If no replacement Ikus is available: Support the patient,
if necessary (in the event of malfunction of the defective Ikus), by means of the
manual pump! (see section 15.5: Driving blood pump(s) with the manual
pump, page 204)
HOTLINE
Notify Berlin Heart! 866.249.0128
14.20 Acoustic alarm is not properly recognized
If the message Acoustic alarm: OK appears within 8 seconds after the error
message,
the Ikus is working perfectly. No further measures are necessary.
If the message Acoustic alarm: OK does not appear within 8 seconds,
the alarm circuit is defective. A possible error might not have been detected.
INSTRUCTION
1.
Check the status of the patient.
2.
Check the filling and emptying of the blood pump(s), and monitor the function
of the Ikus. Do not operate the Ikus without supervision!
3.
Provide the patient immediately with a replacement Ikus and switch off the
malfunctioning Ikus. (see section 15.4: Connecting the patient to a replace
ment Ikus, page 202). If no replacement Ikus is available: Support the patient,
if necessary (in the event of malfunction of the defective Ikus), by means of the
manual pump! (see section 15.5: Driving blood pump(s) with the manual
pump, page 204)
HOTLINE
Notify Berlin Heart! 866.249.0128
14.21 Error: no data/no reaction from the control computer
•
Error: no data from Master.
•
Error: no reaction from Master.
In the event of the simultaneous malfunction of both control computers or the
malfunction of the power supply, the Ikus cannot generate a specific error message in
the message window.
1000721x09 Revision 8
184
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...