Chapter 14 Error Messages and corrective measures
4.
Monitor the membrane movement and make sure that the pump(s) are filling
and ejecting completely. If this is quite correct, confirm the message in the
monitor program.
5.
If the message appears again, then the parameters can be modified slightly to
normalize the flow.
6.
Assess the hemodynamic status of the patient (volumes, MAP, PAP, CVP,
etc.).
7.
Monitor the membrane movement and make sure that the pump(s) are filling
and ejecting completely. If this is functioning properly, then acknowledge the
message in the monitor program.
8.
If the message appears again, check if it is necessary to adjust the cannulae.
If the message appears again:
HOTLINE
Notify Berlin Heart! 866.249.0128
14.5 Backup operation left/right
•
Backup operation (right). Contact costumer service
•
Backup operation (left). Contact costumer service
The left (or right) pneumatic system has failed. The respective pump is now being
powered by the backup system.
INSTRUCTION
1.
Provide the patient immediately with a replacement Ikus.
HOTLINE
Notify Berlin Heart! 866.249.0128
14.6 Error messages in emergency operating mode
14.6.1 UVAD, emergency operation!. Contact service immediately!
UVAD, emergency operation! Contact service immediately!
Appears in univentricular mode.
WARNING
No more system is available as a redundancy. If the only remaining
intact pneumatic system fails, there is a risk that the Ikus will stop
running altogether.
The blood pump is being driven by the last intact pneumatic system.
The pneumatic system continues running with the currently set parameters. The
parameters can still be adjusted if necessary. However, it is not possible to switch over
to biventricular operation.
INSTRUCTION
1.
Provide the patient immediately with a replacement Ikus.
1000721x09 Revision 8
176
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...