Chapter 11 Wound care and treatment
11.2 Cleaning the blood pump
INSTRUCTION
1.
Cleanse the exposed cannula and
the pump head with disinfectant
(i.e. 2% chlorhexidine solution) then
place on sterile towel.
2.
Examine cannulae and cannulae
exit sites.
3.
Remove gloves.
Fig. 11-2
Cleaning the blood pump
Fig. 11-3
Examining the cannulae
11.3 Cleaning of the wound
INSTRUCTION
1.
Hand hygiene, prepare sterile
dressing tray, put on sterile gloves.
If assistance is necessary notify
Berlin Heart.
2.
4X4 gauze soaked in 2% chlorhexi
dine cleanse each cannula exit site
in a circular motion outward to a
radius of approximately 10 cm.
3.
Using a new soaked 4X4 repeat 2
more times beginning at the exit
site and clean in larger circles each
time.
4.
Wrap 4X4 gauze soaked in 2%
chlorhexidine around cannula and
gently cleanse with back/forth
motion.
5.
Repeat with each cannula exit site.
Fig. 11-4
Cleanse each cannula exit site
Fig. 11-5
Cleanse with back/forth motion
1000721x09 Revision 8
137
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...