Chapter 16 Appendix
16.1.10 Maximum rates for the pump-cannula combinations
Cannulation
Blood pumps
Ø inflow
cannula
Ø
outflow
cannula
10 ml
15 ml
25 ml
30 ml
50 ml
60 ml
5 mm
5 mm
130
bpm
6 mm
5 mm
130
bpm
130
bpm
6 mm
6 mm
130
bpm
130
bpm
80 bpm 65 bpm
9 mm
6 mm
130
bpm
100
bpm
90 bpm
9 mm
9 mm
130
bpm
130
bpm
130
bpm
130
bpm
105
bpm
12 mm
9 mm
130
bpm
105
bpm
12 mm
12 mm
130
bpm
125
bpm
Tab. 16-9
Maximum rates for the pump-cannula combinations
Pump-cannula combinations in which not every parameter
combination is recommended (pump rate, % systole, sys
tolic and diastolic pressure) can lead to incomplete filling
and emptying of the blood pump.
The value indicated is the upper threshold for pump rates.
Values that are below the upper threshold are within the
rate value
acceptable range. Values that are higher than the upper
(bpm)
threshold are in a questionable range.
The threshold values have been determinated (in vitro) tak
ing a mean arterial blood pressure of 120 mmHg as a basis.
Red marked values displayed on the laptop: These parame
ter combination (pump rate, % systole, systolic and diastolic
rate value
pressure) for these pump-cannula combination can lead to
(bpm)
incomplete filling and emptying of the blood pump(s).
Observe the filling behavior of the blood pump(s)!
in biventricular mode
The lower value of both pump rates (corresponding to the
pump sizes used) must also be considered. The higher of
the 2 pump rates should be disregarded.
1000721x09 Revision 8
216
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...