Chapter 15 Troubleshooting and correcting faults
The Ikus power
switch (toggle switch) should always be in the [I]
position, even if the main switch (key switch) is in the
[0]
position!.
Otherwise there is a risk that the drive may fail in future due to the Ikus
batteries being totally discharged.
If the graphs in the monitor program are frozen (not moving) and the
parameters cannot be adjusted even after the laptop and the monitor
program have been restarted, then the Ikus is operating in emergency
pulse mode. In this case, do not proceed as described here, but as
described in section 15.3: Emergency pulse mode, page 200 instead.
INSTRUCTION
1.
Support patient with a replacement
Ikus
(see section 15.4: Connecting the
patient to a replacement Ikus, page 202) or manual pump (see section 15.5:
Driving blood pump(s) with the manual pump, page 204).
2.
Use the seal plugs to seal the driving tube connectors.
3.
Switch off the
Ikus
driving unit. In the monitor program, select the
Drive OFF
option and press
<Enter>
to confirm. Respond to the prompt in the dialog win
dow by pressing the
<X>
or the
<1>
key. IMPORTANT: If it is not possible to
select the
Drive OFF
option, the Ikus is running in emergency pulse mode. In
this case, proceed as explained in section 15.3: Emergency pulse mode,
page 200.
4.
Wait until the log has been completed. If the message
Switch off drive with
main switch!
appears, turn the key switch to the
[0]
position.
5.
Switch off the laptop.
6.
Switch Ikus on again immediately. To do so, turn the key switch to the
[I]
posi
tion.
7.
Switch the laptop on. Select the
1. Start program
option (
<1>
). Enter user ID
and password, confirming the password with
<Enter>
.
8.
Check all parameters and re-adjust them if necessary.
9.
For univentricular operation, unplug the connector marked in red, for biventric
ular operation, unplug both connectors. To do so, pull the seal plug(s) out of
the respective connector(s).
10.
Disconnect the driving tube(s) from the replacement
Ikus
or the manual pump
and connect it to the Ikus. IMPORTANT: Observe the colored markings. The
sound of the plug snapping into place is clearly audible.
11.
Check that the plug is securely connected. To do so, grip the plug body above
the release sleeve and pull on it. Do not pull from the release sleeve, and
never from the tube!
12.
Move cursor to the
OK
field and press
<Enter>
to confirm. The standard view
is shown. The system will operate with the current parameter settings.
13.
Switch off the replacement
Ikus
driving unit (see section 6.3.3: Drive OFF:
switching the Ikus off, page 78).
1000721x09 Revision 8
199
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...