Chapter 2 General Information
2.8
Battery replacement and disposal
CAUTION
Only Berlin Heart GmbH/ Berlin Heart, Inc. service staff or persons
authorized by the Berlin Heart GmbH service department may replace
the batteries and dispose of them in accordance with the respective
regulations.
2.9
Manufacturer’s warranty
According to the General Terms and Conditions of Berlin Heart GmbH the warranty is
valid for 1 year.
All warranties apply only under the prescribed conditions of storage of the system, use
in accordance with the instructions (intended use) and when the packaging is intact.
This applies, in particular, to all sterile packaging and to the aluminum-coated outer
packaging of the blood pump(s).
The warranty is no longer valid, if the Ikus has been opened or serviced by persons
who are not members of the Berlin Heart GmbH/ Berlin Heart, Inc. service staff and/
or who have not been authorized by the Berlin Heart GmbH service department to do
so.
If an ambient temperature of +30°C is continuously exceeded during ongoing
operation, the maintenance interval will be reduced.
Components
Product life
(in sterile
products
starting from
implantation)
Maintenance
interval
Expiration
date
Warranty
unsterile
Ikus
max. 8 years
6 months or
2000
operating
hours
x
1 year
Battery
exchange as
needed
6 months or
2000
operating
hours
x
6 months
Manual pump
max. 6 years
yearly
x
1 year
sterile
Driving tube
1 year (single
use product)
x
3 years
1 year
Blood pumps
1 year (single
use product)
x
3 years
1 year
Tab. 2-1
Product life, maintenance interval, expiration date, warranty
1000721x09 Revision 8
23
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...