Chapter 6 Instructions for use: Ikus
6.5.1 Routine start-test when not in operation
IMPORTANT: If the Ikus is not in use, run it once a month for 24 hours in order to
ensure that all batteries are adequately charged. Refer to section 6.5.1: Routine start-
test when not in operation, page 83. Otherwise there is a risk that the Ikus no longer
functions correctly.
INSTRUCTION
1.
Procedure as described in section 7.1.1: Connecting the tank unit, page 94
and section 7.1.2: Switching on the Ikus, page 95.
2.
Wait 24 hours.
3.
Procedure as described in section 6.2.1: Starting the monitor program,
page 71 and section 7.1.4: Setting the test parameters, page 95.
4.
Observe and evaluate the curve representation.
5.
Select the option
Drive OFF
, then confirm with
<Enter>
.
6.
Confirm the selection in the dialog window by pressing
<X>
or
<1>
. The sys
tem immediately stops the operation and writes an operating LOG file.
7.
Wait until the LOG file is complete (message:
Switch off drive with main key
switch
).
8.
End the monitor program and switch off the laptop.
9.
All of the yellow battery LEDs are lighted: Switch off the Ikus. To do so, set the
main switch (key switch) to
[0]
position.
Not all yellow LEDs are lit up: Leave the Ikus on the power supply until all of
the yellow LEDs light up. Then set the main switch (key switch) to
[0]
position.
10.
Disconnect both tank units from the Ikus.
11.
Seal the driving tube connection sockets with seal plugs.
If the system detects an error in the start-up test
INSTRUCTION
1.
End the monitor program according to the instructions in the dialog.
2.
Switch off the Ikus with the main switch (key switch).
IMPORTANT: The power switch (toggle switch) remains on
[I]
position!
3.
IMPORTANT: Wait 10 minutes.
4.
Restart the start-up procedure (see above).
If the system again detects an error in the start-up test or not all of the battery LEDs
are illuminated after 24 hours, notify Berlin Heart:
HOTLINE
Notify Berlin Heart! 866.249.0128
1000721x09 Revision 8
83
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...