Chapter 15 Troubleshooting and correcting faults
15.3.1 Emergency pulse mode - switching the Ikus off
WARNING
Only proceed as described below if a replacement
Ikus
is available to
assume the support of the patient as described here.
The
Ikus
power switch (toggle switch) should always be in the
[I]
position, even if the main switch (key switch) is in the
[0]
position!.
Otherwise there is a risk that the drive may fail in future due to the
Ikus
batteries being totally discharged.
INSTRUCTION
1.
Support the patient by a replacement
Ikus
(see section 15.4: Connecting the
patient to a replacement Ikus, page 202).
2.
Use the seal plugs to seal the driving tube connectors on the
Ikus
.
3.
Switch off the laptop.
4.
Switch off the
Ikus
. To do so, turn the key switch to
[0]
position.
15.3.2 Ikus start-test following emergency pulse mode
WARNING
Do not reconnect the original
Ikus
to the patient until the Berlin Heart,
Inc. service department has evaluated the LOG files or has serviced
the driving unit.
Wait for at least 5 minutes after switching the
Ikus
off while in
emergency pulse mode. Otherwise, only the service staff will be able to
restart it.
Never connect other USB devices (e.g. wireless technology) to the
USB port of the laptop than the delivered USB sticks.
INSTRUCTION
1.
After 5 minutes
: Insert the USB stick into the USB port and switch on the
Ikus
. To do so, turn the key switch to the
[I]
position. The battery charge indica
tor will light up and the number of hours the driving unit has been operated to
date will be displayed. The mains operation indicator lights up.
2.
Switch the laptop on. The menu
Select language
appears.
3.
Select the desired language by pressing the corresponding number key. It is
not necessary to press
<Enter>
to confirm this selection. The start menu is
displayed on the laptop.
4.
Select the
1. Start program
option (
<1>
). Enter user ID and password, con
firming the password with
<Enter>
. The
Ikus
will carry out a start-test.
5.
Wait for the start-test phase to finish (this takes a few minutes). Do not mute
the acoustic signal. The messages in the message window inform the user of
the current status of the test. If the system is found to be operating correctly,
the view
Select operating mode
will be displayed next.
6.
Select
Drive OFF
, then press
<Enter>
to confirm.
1000721x09 Revision 8
201
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...