Appendix C - Compressor Controller (Mk5)
C.5.1 Regulation
Name
Description
Controller
Default Value
Beacon
Medaes Value
Unloading Pressure 1
High pressure level of Pressure band 1
8 bar
105 psi¹
Loading Pressure 1
Low pressure level of Pressure band 1
6.4 bar
96 psi¹
Unloading Pressure 2
High pressure level of Pressure band 1
Loading Pressure 2
Low pressure level of Pressure band 1
Pressure Band Used
1
1
Digital Pressure Selection Band
Off
Off
Minimum Stop Time
Time which the compressor has to stop
before restarting
20 seconds
20 seconds
Unload Time
30 seconds
30 seconds
Start check time
20 seconds
20 seconds
Automatic Restart²
Not activated
Infinite²
Maximum Power Down
15 seconds
15 seconds
Restart
0 seconds
0 seconds
Ethernet
On
On
CAN³
CAN must be off to change CAN
address and communication profile
Off
On³
CAN Address
31
Node ID
Communication Profile
Mk4
Mk4
1. Typical settings shown for lead compressor. Lower these settings by 5 psi for each remaining
compressor so that all compressors have a different local set point.
2. Password required to set “Auto Restart” parameter to infinite (4735)
3. CAN must be off to change CAN address and communication profile.