E-2
Z MED Medical Air System
4107 9022 67.00
Appendix E - Dewpoint Transmitter
E.5 Alarms
Dewpoints that exceed setpoints shall cause
an alarm condition at the Central Control Panel
(CCP). When the dew point exceeds the alarm
set point, the CCP contacts are de-energized.
The alarm remains de-energized until the alarm
condition is cleared by the operator.
E.6 Maintenance
E.6.1 Repair Policy
Do not use a unit that is not functioning
properly until all necessary repairs have
been made and the unit has been tested
to determine that it is functioning in
accordance with the manufacturer’s published
specifications. Contact
BeaconMedæs
Technical
Services department at 1-888-4MEDGAS (888-
463-3427) for assistance.
E.6.2 Maintenance Schedule
WARNING!
Respond to alarm conditions
immediately. An alarm condition on
the dewpoint indicates a dew point
level exceeding the maximum set point.
Immediate action to correct the problem
should be taken. Prolonged exposure to
condensing moisture can damage the
Medical Air equipment.
Maintenance
Frequency
Action
Check flow through flowmeter
Weekly
Check for proper flow (1.0 - 2.0 SCFH)
Check Transmitter accuracy
Every 12 months
Verify dewpoint sensor accuracy (contact
BeaconMedæs
)
Recalibration
Every 2 years
See Section E.8
Replace sensor
Every 5 years
See Section E.9
WARNING!
* Electrical shock hazard *
No repair should ever be attempted by
anyone not having experience in the
repair of devices of this nature. Failure
to follow proper repair procedures can
result in serious injury.
Note:
To ensure full reliability, have all
maintenance and testing done by a qualified
technician. If this cannot be done, maintenance
and testing of those parts discussed in this
manual may be undertaken by a competent,
trained individual having experience in the
repair of devices of this nature
CAUTION:
No maintenance and testing
should ever be undertaken or attempted by
anyone not having general experience in the
repair of devices of this nature. Also, to avoid
damaging the unit or any of its components, no
maintenance and testing should be undertaken
by qualified individuals who are not familiar
with the procedures in this manual.
Replace damaged parts with components from
BeaconMedæs. Test the unit after installation
of replacement parts to make certain that it
complies with the published specifications.