A-10
Z MED Medical Air System
4107 9022 67.00
Name
Description
Controller
Default Value
Beacon
Medaes Value
‘C.C.M.’
ES6 Control Mode (Local, Remote, LAN)
Local
Local
‘Pressure Band Selec’ Allows manual selection between pressure band 1 or 2
Press Band 1
Press Band 1
‘Digital PB sel.’
When this function is activated the pressure band will
be controlled by a switch (connected to a digital input).
Once this function is activated, the switch position over-
rules the selection in the menu.
Not Activated
Not Activated
A.6.1.2 Configuration Listing
See A.6.2.10 (Modify Parameters – Configuration) for screen navigation to view parameter settings.
Name
Description
Controller
Default Value
Beacon
Medaes Value
‘S1 ID1 Priority’
Scheme 1. Priority level for connection channel 1
1
1
‘S1 ID2 Priority’
Scheme 1. Priority level for connection channel 2
1
1
‘S1 ID3 Priority’
Scheme 1. Priority level for connection channel 3
1
1
‘S1 ID4 Priority’
Scheme 1. Priority level for connection channel 4
1
1
‘S1 ID5 Priority’
Scheme 1. Priority level for connection channel 5
1
1
‘S1 ID6 Priority’
Scheme 1. Priority level for connection channel 6
1
1
‘S2 ID1 Priority’
Scheme 2. Priority level for connection channel 1
1
1
‘S2 ID2 Priority’
Scheme 2. Priority level for connection channel 2
1
1
‘S2 ID3 Priority’
Scheme 2. Priority level for connection channel 3
1
1
‘S2 ID4 Priority’
Scheme 2. Priority level for connection channel 4
1
1
‘S2 ID5 Priority’
Scheme 2. Priority level for connection channel 5
1
1
‘S2 ID6 Priority’
Scheme 2. Priority level for connection channel 6
1
1
‘Running Hours ID1’
Scheme 2. Priority level for connection channel 1
0
0
‘Running Hours ID2’
Scheme 2. Priority level for connection channel 2
0
0
‘Running Hours ID3’
Scheme 2. Priority level for connection channel 3
0
0
‘Running Hours ID4’
Scheme 2. Priority level for connection channel 4
0
0
‘Running Hours ID5’
Scheme 2. Priority level for connection channel 5
0
0
‘Running Hours ID6’
Scheme 2. Priority level for connection channel 6
0
0
‘Power Recovery
Time’
Maximum allowed power down time to accept the Auto-
matic Restart After Voltage Failure (ARAVF)
10 seconds
oo! sec
*
‘Restart Delay’
Start delay after the voltage is restored
5 seconds
5 seconds
Communication
Time-Out
**
Delay time prior to activating a communication general
fault alarm, when communication to connected machines
has been lost or not re-established after power outage.
5 seconds
30 seconds
* Controller display will show “oo! sec” when “Auto Restart” parameter is set to infinite
** Can only be updated using a laptop and FSP+ Software
Appendix A - Central Control System
A.6.1.1 Parameter Listing (Continued)